Specifications
B-3
CAL 8000 (without
pneumatic unit)
Varies with
specific
configurations
Varies with
specific
configurations
Nonterminal analyzer track
module
Nonterminal SM&S track
module
B.10 Contraindication
None
B.11 Barcode Specifications
The Sample Processing System can read barcodes that are stuck to the test tubes. The
sample ID read from the barcode will be stored and used as the only identification of the
sample.
The barcodes used shall meet the specifications stated in this section.
1. Supported barcode types
All code types and check digit supported by the system are listed as follows.
Barcodes Supported by the Sample Processing System
Self-checking (check digit is
always included)
No more than 20 digits (sample ID)
Self-checking (check digit is
always included)
No more than 20 digits (sample ID)
Self-checking (check digit is
always included)
Fixed length: 8 or 13 digits