10-2 Passport 12/Passport 8 Operators Manual
10.3 Selecting the Active Alarm Source
In most cases the HR and pulse numerics are identical. In order to avoid simultaneous alarms on HR and PR, the monitor
uses either HR or PR as its active alarm source. To change the alarm source, select [Alm Source] in the [ECG Setup] or
[SpO2 Setup] menu and then select either:
[HR]: The monitor will use the HR as the alarm source for HR/pulse.
[PR]: The monitor will use the PR as the alarm source for HR/pulse.
[Auto]: If the [Alm Source] is set to [Auto], the monitor will use the heart rate from the ECG measurement as the
alarm source whenever the ECG measurement is switched on and a valid heart rate is available. If the heart rate
becomes unavailable, for example if leads becomes disconnected, and a pulse source is switch on and available, the
monitor will automatically switch to Pulse as the alarm source. When the Leads Off condition is corrected, the
monitor will automatically switch back to the heart rate as the alarm source.
10.4 QRS Tone
When PR is used as the alarm source, the PR source will be used as a source for the QRS tone. You can change the QRS
volume by adjusting [Beat Vol] in the [SpO2 Setup] menu. When a valid SpO
2
value exists, the system will adjust the
pitch tone of QRS volume according to the SpO
2
value.

Table of Contents

Other manuals for Mindray Passport 12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 12 and is the answer not in the manual?

Mindray Passport 12 Specifications

General IconGeneral
BrandMindray
ModelPassport 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals