7-4 Passport 12/Passport 8 Operators Manual
7.4 Alarm Tone Configuration
7.4.1 Setting the Minimum Alarm Volume
To set the minimum alarm volume:
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Minimum Alarm Volume] and then select a value between 0 and 10.
The minimum alarm volume refers to the minimum value you can set for the alarm volume, which is not affected by user
or factory default configurations. The minimum alarm volume setting remains unchanged when the monitor shuts
down and restarts.
7.4.2 Changing the Alarm Volume
To change the alarm volume:
1. Select the [Volume Setup] QuickKey or the [Alarm Setup] QuickKey[Others], or [Main Menu][Alarm
Setup >>][Others].
2. Select the appropriate volume from [Alm Volume]: X to 10, in which X is the minimum volume, depending on the
set minimum alarm volume, and 10 the maximum volume.
3. Select [High Alarm Volume] to set the high priority alarm volume as [Alm Volume+0], [Alm Volume+1] or [Alm
Volume+2].
4. Select [Reminder Vol] to set the volume of the reminder tone as [High], [Med] or [Low].
When alarm volume is set to 0, the alarm sound is turned off and a
symbol appears in the alarm symbols area of the
screen.
7.4.3 Setting the Interval Between Alarm Sounds
You cannot change the interval between alarm tones if you choose mode 1 or 2 as your desired alarm tone pattern. For
these two patterns, the interval between alarm tones identifies the alarm levels as follows:
Mode 1:
Interval between high level alarm tones: continuously
Interval between medium level alarm tones: 5 s
Interval between low level alarm tones: 20 s
Mode 2:
Interval between high level alarm tones: 1 s
Interval between medium level alarm tones: 5 s
Interval between low level alarm tones: 20 s

Table of Contents

Other manuals for Mindray Passport 12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 12 and is the answer not in the manual?

Mindray Passport 12 Specifications

General IconGeneral
BrandMindray
ModelPassport 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals