EasyManuals Logo
Home>Mindray>Medical Equipment>Passport 12

Mindray Passport 12 Operator's Manual

Mindray Passport 12
284 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #155 background imageLoading...
Page #155 background image
Passport 12/Passport 8 Operators Manual 16-9
16.10 Removing Exhaust Gases from the System
WARNING
Anesthetics: When using the Sidestream or Microstream CO
2
measurement on patients who are receiving or
have recently received anesthetics, connect the outlet to a scavenging system, or to the anesthesia
machine/ventilator, to avoid exposing medical staff to anesthetics.
To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet connector of the module.
16.11 Zeroing the Sensor
Zero calibration eliminates the effect of baseline drift during CO
2
measurement exerted on the readings and therefore
maintains the accuracy of the CO
2
measurements.
For sidestream and microstream CO
2
modules, a zero calibration is carried out automatically when necessary. You can
also start a manual zero calibration if necessary.
To manually start a zero calibration:
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the required password[Maintain
CO2 >>].
2. Select [Calibrate CO2 >>][Start Zero Cal.].
Disconnecting the patient airway is not required when performing a zero calibration.
WARNING
Do not rely on the readings during zeroing.

Table of Contents

Other manuals for Mindray Passport 12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 12 and is the answer not in the manual?

Mindray Passport 12 Specifications

General IconGeneral
BrandMindray
ModelPassport 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals