Passport 12/Passport 8 Operators Manual 7-9
Module Parameter
Low alarm limit High alarm limit
Auto alarm limits
range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate
AG
EtCO
2
(AG)
Same as CO
2
module
FiCO
2
(AG)
awRR
awRR × 0.5 or 6
rpm (whichever is
greater)
awRR – 10 or 30
rpm (whichever is
greater)
awRR × 1.5 or 30
rpm (whichever is
smaller)
awRR + 25 or 85
rpm (whichever is
smaller)
Adult/pediatric: 6 to
55
Neonate: 10 to 90
FiAA/
EtAA
Same as the
default alarm limit
Same as the
default alarm limit
Same as the
default alarm limit
Same as the default
alarm limit
Same as the
measurement range
FiO
2
/
EtCO
2
Same as the
default alarm limit
Same as the
default alarm limit
Same as the
default alarm limit
Same as the default
alarm limit
Same as the
measurement range
FiN
2
O/
EtN
2
O
Same as the
default alarm limit
Same as the
default alarm limit
Same as the
default alarm limit
Same as the default
alarm limit
Same as the
measurement range
C.O. BT
Adult:
(BT – 1)°C
N/A
Adult:
(BT – 1)°C
N/A
Same as the
measurement range
7.5.3 Setting Alarm Delay Time
You can set the alarm delay time for alarms of continuously measured parameters. If the alarm condition is resolved
within the delay time, the monitor will not sound the alarm.
To set the alarm delay time:
1. Select [Main Menu][Maintenance >>][User Maintenance >>]. Enter the required password and then select
[Ok].
2. Select [Alarm Setup >>][Alarm Delay].
Alarm delay is not applied to the following physiological alarms:
Apnea
ST alarms
Arrhythmia alarms
ECG Weak Signal
Resp Artifact
SpO
2
Desat
No Pulse
Nellcor SpO
2
over alarm limits
FiO
2
Shortage
Measurements of noncontinuous parameters over alarm limits
HR over alarm limits
Anesthetic Mixture's MAC > 3
You can set [Apnea Delay] and [ST Alarm Delay] separately.

Table of Contents

Other manuals for Mindray Passport 12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 12 and is the answer not in the manual?

Mindray Passport 12 Specifications

General IconGeneral
BrandMindray
ModelPassport 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals