7-8 Passport 12/Passport 8 Operator’s Manual
Module Parameter
Low alarm limit High alarm limit
Auto alarm limits
range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate
Temp
T1 (T1 – 0.5)°C (T1 – 0.5) °C (T1 + 0.5)°C (T1 + 0.5)°C 1 to 49°C
T2 (T2 – 0.5)°C (T2 – 0.5) °C (T2 + 0.5)°C (T2 + 0.5)°C 1 to 49°C
TD
Same as the
default alarm limit
Same as the
default alarm limit
Same as the
default alarm limit
Same as the default
alarm limit
Same as the
measurement range
IBP: ART/
Ao/
UAP/
BAP/
FAP/
LV/
P1-P4
(Arterial
pressure)
IBP-S
(SYS × 0.68+10)
mmHg
(SYS – 15) or 45
mmHg (whichever
is greater)
(SYS × 0.86 + 38)
mmHg
(SYS + 15) or 105
mmHg (whichever
is smaller)
Adult: 45 to 270
Pediatric: 45 to 185
Neonate: 35 to 115
IBP-D
(Dia × 0.68 + 6)
mmHg
(Dia – 15) or 20
mmHg (whichever
is greater)
(Dia × 0.86 + 32)
mmHg
(Dia + 15) or 80
mmHg (whichever
is smaller)
Adult: 25 to 225
Pediatric: 25 to 150
Neonate: 20 to 90
IBP-M
(Mean × 0.68 + 8)
mmHg
(Mean – 15) or 35
mmHg (whichever
is greater)
(Mean × 0.86 +
35) mmHg
(Mean + 15) or 95
mmHg (whichever
is smaller)
Adult: 30 to 245
Pediatric: 30 to 180
Neonate: 25 to 105
PA
IBP-S SYS × 0.75 SYS × 0.75 SYS × 1.25 SYS × 1.25
3 to 120 mmHg IBP-D Dia × 0.75 Dia × 0.75 Dia × 1.25 Dia × 1.25
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25
IBP:
CVP/
ICP/
LAP/
RAP/
UVP/
P1-P4
(Venous
pressure)
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40mmHg
CO
2
EtCO
2
0 to 32 mmHg:
remains the same
0 to 32 mmHg:
remains the same
0 to 32 mmHg:
remains the same
0 to 32 mmHg:
remains the same
Same as the
measurement range
32 to 35 mmHg:
29 mmHg
32 to 35 mmHg:
29 mmHg
32 to 35 mmHg:
41 mmHg
32 to 35 mmHg:
41 mmHg
35 to 45 mmHg:
(etCO
2
-6) mmHg
35 to 45 mmHg:
(etCO
2
-6) mmHg
35 to 45 mmHg:
(etCO
2
+6) mmHg
35 to 45 mmHg:
(etCO
2
+6) mmHg
45 to 48 mmHg:
39 mmHg
45 to 48 mmHg:
39 mmHg
45 to 48 mmHg:
51 mmHg
45 to 48 mmHg:
51 mmHg
> 48 mmHg:
remains the same
> 48 mmHg:
remains the same
> 48 mmHg:
remains the same
> 48 mmHg:
remains the same
FiCO
2
N/A N/A
Same as the
default alarm limit
Same as the default
alarm limit
Same as the
measurement range
awRR
awRR × 0.5 or 6
rpm (whichever is
greater)
(awRR – 10) or 30
rpm (whichever is
greater)
awRR × 1.5 or 30
rpm (whichever is
smaller)
(awRR + 25) or 85
rpm (whichever is
smaller)
Adult/pediatric: 6 to
55
Neonate:10 to 90

Table of Contents

Other manuals for Mindray Passport 12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 12 and is the answer not in the manual?

Mindray Passport 12 Specifications

General IconGeneral
BrandMindray
ModelPassport 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals