P/N: H-046-005847-00 Rev 3.0
6
Passport
®
Quick Reference Guide
Or
a. Select ECG Setup>> by selecting ECG waveform/ HR numeric tile or from
Parameters menu ( or Main Menu quick keys)
b. Select ECG to adjust lead (3 lead) or select ECG1 to adjust first lead and
ECG2 to adjust the second lead (5 lead)
c. Scroll and select lead from drop down menu
d. Select Main Menu (quick key) or to return to the main display
DPM Central Station Interface: Adjusting ECG leads on the Passport 8 or
Passport 12 bedside monitor and DPM Central Station are independent.
Panorama Central Station Interface: Adjusting ECG leads on the Passport 8
or Passport 12 bedside monitor, Panorama Central Station, WorkStation, and
ViewStation are independent.
Enable Pacemaker Detection And Set Pacemaker Rate
The Passport 8 or Passport 12 bedside monitors allow the clinician to enable
pacemaker detection. The symbol is displayed when the pacemaker
detection is enabled.
To enable the pacemaker detection on the Passport 8 or Passport 12 bedside
monitor:
a. Select ECG waveform or ECG numeric tile to enter the ECG Setup menu
Or
a. Select ECG Setup>> by selecting ECG waveform/ HR numeric tile or from
Parameters menu ( or Main Menu quick keys)
b. Select Others >>
c. Select Paced to Yes
d. Select Pacemaker Rate value and adjust using the up and down arrows
e. Select Main Menu (quick key) or to return to the main display
DPM Central Station Interface: Pacemaker detection on the Passport 8 or
Passport 12 bedside monitor and DPM Central Station are independent.