P/N: H-046-005847-00 Rev 3.0
14
Passport
®
Quick Reference Guide
Station, WorkStation, and ViewStation for a preconfigured amount of time.
Selecting Mute on the Panorama Central Station will not silence an alarm on
the Passport 8 or Passport 12 bedside monitor.
SILENCE ALL ALARMS
To silence all alarms on the Passport 8 or Passport 12 bedside monitor for a
preconfigured amount of time, select Alarm Pause key on front panel of
bedside monitor.
DMP Central Station Interface: Selecting Alarm Pause on the Passport 8
or Passport 12 bedside monitor will silence active alarms on the Panorama
Central Station, WorkStation, and ViewStation for a preconfigured amount of
time. Selecting Alarm Pause on the Panorama Central Station will silence
alarms on the Passport 8 or Passport 12 bedside monitor.
Panorama Central Station Interface: Selecting Alarm Pause on the Passport
8 or Passport 12 bedside monitor will silence active alarms on the Panorama
Central Station, WorkStation, and ViewStation for a preconfigured amount of
time. Selecting Mute on the Panorama Central Station will not silence alarms
on the Passport 8 or Passport 12 bedside monitor.
TRANSFER*
The Passport 8 or Passport 12 bedside monitors have the ability to transfer
patient data through a storage medium (USB Flash Drive)
To transfer patient data from one Passport 8 or Passport 12 bedside monitor to
another Passport 8 or Passport 12 bedside monitor:
a. Insert USB Flash Drive into USB port on back of monitor
b. Select Patient Setup >> using Patient Setup (quick key) or Main
Menu (quick key)
c. Select Transfer to Storage Medium*
d. Select OK
e. Remove USB Flash Drive once Transfer to storage medium successful
message is displayed
f. Insert USB Flash Drive into USB port on back of new Passport 8 or
Passport 12 bedside monitor