Panorama Exposures
22
(3) Panorama Settings
When the unit is turned on, the operation panel is set to the
factory defaults shown in the photo to the left.
(3)-1. Auto Exposure (Digital Direct Auto Exposure)
For auto exposure, the X-ray dosage is monitored and adjusted
in real time depending on the patient’s anatomy and the
selected exposure region.
Press the Auto Level key to use auto exposure.
The current auto level value is shown on the left side of the
Auto Level key. This value can be changed. Up and Down
keys will also appear to the right; use these to change the
settings.
Auto level can be adjusted from +4 to -4 in increments of 1.
* The factory setting for the Auto Level is 0.
Chart 1 shows the relative X-ray output for each auto level.
Chart 1
Auto Level
X-ray Output Ratio
+4
286%
+3
220%
+2
169%
+1
130%
0
100%
-1
77%
-2
59%
-3
46%
-4
35%