Panorama Exposures
(3)-2. Manual Exposure
Press the kV or mA key to switch to manual exposure.
Press the kV key to display the tube voltage value. This value
can be changed. Use the Up and Down keys to increase or
decrease the value.
Press the mA key to display the tube current value. This value
can be changed. Use the Up and Down keys to increase or
decrease the value.
kV can be set from 60 to 80 kV in increments of 1 kV, and mA
from 1 to 10 mA in increments of 1 mA.
* The factory settings are 65 kV and 5 mA.
(3)-3. No X-ray Setting
Use this to rotate the arm to make sure it will not hit the
patient or for other tests which do not require X-ray emission.
To activate No X-ray mode, hold down the Down key until the
Auto Level, kV and mA displays go out.
To restore normal operation, hold down the Up key.
To use the No X-ray mode, make sure the Auto Level,
kV and mA displays are blank.