Which Method Is Best?
This depends on the frequency to be measured, the rate at which you want to monitor the
frequency and the accuracy you desire. Take for example, measuring a 50 kHz signal. Assuming
that the measurement times for the sample clocked (with averaging) and two counter frequency
measurements are configured the same, Table 6-3 summarizes the results.
From this, you can see that while the measurement time for one counter is shorter, the accuracy
is best in the sample clocked and two counter large range measurements. For another example,
Table 6-4 shows the results for 5 MHz.
Table 6-3. 50 kHz Frequency Measurement Methods
Variable
Sample
Clocked
One Counter
Two Counters
High
Frequency
Large Range
fx 50,000 50,000 50,000 50,000
fk 80 M 80 M 1,000 80 M
Measurement
time (mS)
1 .02 1 1
N 50
Max. frequency
error (Hz)
.638 31.27 1,000 .625
Max. error % .00128 .0625 2 .00125
Table 6-4. 5 MHz Frequency Measurement Methods
Variable Sample Clocked One Counter
Two Counters
High
Frequency
Large Range
fx 5 M 5 M 5 M 5 M
fk 80 M 80 M 1,000 80 M
Measurement
time (mS)
1 .0002 1 1
N 5,000
Max.
Frequency
error (Hz)
62.51 333 k 1,000 62.50
Max. Error % .00125 6.67 .02 .00125
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments cDAQ-9185 and is the answer not in the manual?

National Instruments cDAQ-9185 Specifications

General IconGeneral
BrandNational Instruments
ModelcDAQ-9185
CategoryChassis
LanguageEnglish

Related product manuals