© National Instruments | 6-9
cDAQ-9185/9189 User Manual
Figure 6-9 shows an example of an implicit buffered pulse measurement.
Figure 6-9. Implicit Buffered Pulse Measurement
Sample Clocked Buffered Pulse Measurement
A sample clocked buffered pulse measurement is similar to single pulse measurement, but a
buffered pulse measurement takes measurements over multiple pulses correlated to a sample
clock.
The counter performs a pulse measurement on the Gate. On each sample clock edge, the counter
stores the high and low ticks in the FIFO of the last pulse to complete. The STC3 transfers the
sampled values to host memory using a high-speed data stream.
Figure 6-10 shows an example of a sample clocked buffered pulse measurement.
Figure 6-10. Sample Clocked Buffered Pulse Measurement
Note If a pulse does not occur between sample clocks, an overrun error occurs.
For information about connecting counter signals, refer to the Default Counter/Timer Routing
section.
Gate
Source
HL
42
HL
4
2
44
6
2
2
2
HL
4
2
4
4
6
2
HL
42
4
4
Buffer
Counter Armed
Counter
Armed
Gate
Source
HL
2
2
HL
22
3
3
Sample
Clock
S1
S2
Buffer
2
2
3
3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments cDAQ-9185 and is the answer not in the manual?

National Instruments cDAQ-9185 Specifications

General IconGeneral
BrandNational Instruments
ModelcDAQ-9185
CategoryChassis
LanguageEnglish

Related product manuals