DB Series Nitrogen Generator DB5-20 Model
5.4 Configuring SW2400-MOD MODBUS-TCP to MODBUS-RTU Option
Before proceeding, set the IP settings in the dryer to the following:
Refer to CONFIGURING STATIC IP ADDRESS
Mode: STATIC
IP Address: 19 2.168 .1.2 0 0
Subnet Mask: 255.255.255.0
Gateway: 0.0.0.0
DNS: 0.0.0.0
The software that comes with the gateway is used to change the configuration
1. Install the MGate Manager software included on the CD that came with the gateway
2. Connect the Gateway to Ethernet and power up the device
3. Select Search and choose Broadcast Search to find the device on the network
4. Highlight the device and select Configuration