Page 186
Chapter 9 Troubleshooting
This section describes many common symptoms, with suggestions as to
what to do.
Symptom
What to Do
Display blank, transducer not firing.
Check power supply, voltage
selector switch and fuse.
Displays “No Xducer”
Check wiring to transducer.
Displays “Xducer Flt”
There is a fault with the
transducer wiring, so check
wiring to transducer.
Incorrect reading being displayed for
current level.
Measure actual distance from
transducer head to surface of
material. Enter Program Mode
and directly access P21 (Set
Distance) type in the measured
distance, ENTER, ENTER
again when prompted, wait
until SET displayed and return
to Run Mode, display should
now update to correct reading.
Material level is consistently incorrect
by the same amount.
Check empty level, (P105)
display offset, (P802) and
measurement offset (P851).
LED’s change colour at relevant relay
switch points but relays do not change
state.
Check supply to unit and
ensure voltage selector set to
correct position.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pulsar Ultra 3 (UL) and is the answer not in the manual?

Pulsar Ultra 3 (UL) Specifications

General IconGeneral
BrandPulsar
ModelUltra 3 (UL)
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals