- 91 -
25. ERROR MESSAGES
Message
Error
code
Error description
"Control sum error."
1.1
Error related to data transmission
" A/D converter error"
1.2
Converter error
"Measuring range
exceeded"
2.1
Maximum capacity (measuring range) of the
balance exceeded (over load)
"Measuring range
exceeded"
2.2
Maximum capacity (measuring range) of the
balance exceeded (over load)
"A/D Null"
2.3
No divisions from the converter
"A/D Full"
2.4
Maximum quantity of divisions from
converter exceeded
"Tare/Zero out of range"
2.5
Permissible value of zeroing or tarring range
exceeded
"Tare out of range"
2.6
Permissible value of tarring range exceeded
"Zero out of range”
2.7
Permissible value of zeroing range
exceeded
"Result > 4% Max"
2.8
Too high start mass (balance started with a
load on the weighing pan)
"Result > 1% Max"
2.9
The difference between mass of a currently
measured adjustment weight and mass
saved in balance memory is greater than
(difference >1%)
"Part < 1 Div"
2.10
Mass value of a single part in Parts Counting
mode is smaller than balance reading unit
“Part < 10 Div"
2.11
Load value placed on the weighing pan
while determining mass a single part in Parts
Counting mode is smaller than 10 reading
units of a balance
"Ref < 1000 Div"
2.12
Value of a reference mass in Percent Setup
mode in smaller than 1000 reading units
"Out of range"
3.1
Parameter value exceeds permissible value
"Incorrect value "
3.2
Impermissible parameter value
"DRH - locked"
3.3
Parameter change blocked (DRH function in
the factory menu is enabled)
"Overwriting error "
4.1
Errors relating to data transmission to a
computer or a printer connected to a
balance
"Parity error"
4.2
"Frame error"
4.3
"CTS transmission
suspended "
4.4
"XOFF transmission
suspended "
4.5
"Incorrect date"
5.1
Error of date setting
"Time value exceeded"
6.1
Timeout error for carrying out a given
process (e.g. zeroing).

Table of Contents