Remote Control
R&S
®
FPC
277User Manual 1178.4130.02 ─ 07
<m> Marker
Example: //Match the reference level to the level of marker 1
CALC:MARK1:FUNC:REF
Usage: Event
Manual operation: See "Center = Marker / Level = Marker" on page 376
See "Center = Marker / Level = Marker" on page 386
See "Center = Marker / Level = Marker" on page 404
CALCulate<n>:MARKer<m>:FUNCtion:STRack[:STATe] <State>
This command turns marker tracking on and off.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Turn on marker tracking
CALC:MARK:FUNC:STR ON
Manual operation: See "Marker Tracking" on page 377
See "Marker Tracking" on page 386
See "Marker Tracking" on page 404
CALCulate<n>:MARKer<m>:MAXimum:NEXT
This command moves a marker to the next higher trace maximum.
If the marker is not active yet, the command also turns on that marker.
Suffix:
<n>
.
irrelevant
<m> Marker
Example: //Move marker 2 to the next higher value
CALC:MARK2:MAX:NEXT
Usage: Event
Manual operation: See "Set To Next Peak" on page 376
See "Set To Next Peak" on page 386
See "Set To Next Peak" on page 404
CALCulate<n>:MARKer<m>:MAXimum[:PEAK]
This command positions a marker on the current trace maximum.
If the marker is not active yet, the command also turns on that marker.
Remote Commands of the Spectrum Application