Remote Control
R&S
®
FPC
281User Manual 1178.4130.02 ─ 07
If you perform a measurement in zero span, this command limits the evaluation range
on the trace.
Suffix:
<n>, <m>, <slim>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF | 0
Example: See CALCulate<n>:MARKer<m>:X:SLIMits<slim>:RIGHt
Manual operation: See "Search Range" on page 377
See "Search Range" on page 387
See "Search Range" on page 405
CALCulate<n>:MARKer<m>:Y?
This command queries the absolute position of a marker on the y-axis.
If necessary, the corresponding marker is activated first.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
Suffix:
<n>
.
irrelevant
<m> Marker
Return values:
<Position> <numeric value>
The unit depends on UNIT<n>:POWer
Example: //Turn on single sweep mode and marker 2
INIT:CONT OFF
CALC:MARK2 ON
//Start a sweep and wait for the end
INIT;*WAI
//Query the position of marker 2
CALC:MARK2:X?
CALC:MARK2:Y?
Usage: Query only
Manual operation: See "Marker <x>" on page 374
See "Marker <x>" on page 384
See "Marker <x>" on page 395
Remote Commands of the Spectrum Application