Remote Control
R&S
®
FPC
312User Manual 1178.4130.02 ─ 07
Return values:
<Time> <numeric value>
Default unit: s
Example: //Query measurement time
ADEM:MTIM?
Usage: Query only
19.14.6 Limits
Commands to configure limits described elsewhere.
●
Beeper:
CALCulate<n>:LIMit<k>:BEEP[:STATe]
CALCulate<n>:ADEMod:LIMit:DELete..............................................................................312
CALCulate<n>:ADEMod:LIMit:FAIL?................................................................................312
CALCulate<n>:ADEMod:LIMit:SELect..............................................................................313
CALCulate<n>:ADEMod:LIMit:DELete
This command removes a set of limit values from the measurement. Limit check is no
longer possible.
Prerequisites for this command
●
Load a set of limit values (CALCulate<n>:ADEMod:LIMit:SELect).
Suffix:
<n>
.
irrelevant
Example: //Delete limit values
CALC:ADEM:LIM:DEL
Usage: Event
Manual operation: See "Clear Limits" on page 390
CALCulate<n>:ADEMod:LIMit:FAIL?
This command queries the results of the limit check.
Prerequisites for this command
●
Load a set of limit values (CALCulate<n>:ADEMod:LIMit:SELect).
Suffix:
<n>
.
irrelevant
Example: //Query limit check results
CALC:ADEM:LIM:FAIL?
Usage: Query only
Remote Commands of the Analog Demodulation