RIGOL Chapter 2 Command System
2-180 MSO1000Z/DS1000Z Programming Guide
:TRIGger:DURATion:TYPe
Syntax
:TRIGger:DURATion:TYPe <type_ch1>[,<type_ch2>[,<type_ch3>
[,<type_ch4>[,<type_d0>…[,<type_d15>]]]]]
:TRIGger:DURATion:TYPe?
Description
Set or query the pattern of each channel in duration trigger.
Parameter
Name
Type
Range
Default
<type_ch1>
Discrete
{H|L|X}
X
<type_ch2>
Discrete
{H|L|X}
X
<type_ch3>
Discrete
{H|L|X}
X
<type_ch4>
Discrete
{H|L|X}
X
<type_d0>
Discrete
{H|L|X}
X
……
……
……
……
<type_d15>
Discrete
{H|L|X}
X
Explanation
 <type_ch1> to <type_ch4> set the patterns of analog channels CH1 to CH4.
<type_d0> to <type_d15> set the patterns of digital channels D0 to D15.
 Users can send 20 parameters to set the pataterns of all the channels. Users can also
omit some parameters to set the patterns of some of the channels (the pattern states
of the channels of which the parameters are omitted remains unchanged), but at
least one parameter (this parameter sets the pattern of CH1) should be sent. When
the number of parameters sent is less than 20, the instrument sets the channels in
CH1 to CH4 and D0 to D15 order.
 In the range of the parameter, H represents high level (higher than the threshold
level of the channel). L represents low level (lower than the threshold level of the
channel). X denotes that this channel is ignored (this channel is not used as a part of
the pattern. When all the channels in the pattern are set to X, the oscilloscope will not
trigger).
 Please distinguish "omit parameter" from "ignore channel". The former means that
the parameters corresponding to some channels are not sent when sending the
command and the pattern states of these channels remain unchanged. The latter
means that the channel is set to X when sending the command and the state of this
channel is ignored when the instrument judges the patterns.
 The query returns the patterns of the 4 analog channels when all the digital channels
are turned off; the query returns the patterns of all the 20 channels when any of the
digital channels is turned on.
Return
Format
The query returns the patterns of the 4 analog channels or all the channels. The patterns
of multiple channels are separated by commas.
Example
:TRIGger:DURATion:TYPe L,X,H,L /*Set the patterns of CH1 to CH4 to L,X,H,L. The
patterns of the other channels remain unchanged*/
:TRIGger:DURATion:TYPe? /*The query returns
L,X,H,L,X,X,X,X,X,X,X,X,X,X,X,X,X,X,X,X*/

Table of Contents

Other manuals for Rigol DS1104Z-S Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z-S Plus and is the answer not in the manual?

Rigol DS1104Z-S Plus Specifications

General IconGeneral
BrandRigol
ModelDS1104Z-S Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals