Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-181
:TRIGger:DURATion:WHEN
Syntax
:TRIGger:DURATion:WHEN <when>
:TRIGger:DURATion:WHEN?
Description
Set or query the trigger condition in duration trigger.
Parameter
Name
Type
Range
Default
<when>
Discrete
{GREater|LESS|GLESs}
GREater
Explanation
GREater: you need to specify a time (refer to :TRIGger:DURATion:TLOWer). The
oscilloscope triggers when the duration of the pattern is greater than the preset time.
LESS: you need to specify a time (refer to :TRIGger:DURATion:TUPPer). The
oscilloscope triggers when the duration of the pattern is lower than the preset time.
GLESs: you need to specify an upper limit of the time (refer
to :TRIGger:DURATion:TUPPer) and lower limit of the time (refer
to :TRIGger:DURATion:TLOWer). The oscilloscope triggers when the duration of the
pattern is lower than the preset upper limit of the time and greater than the preset
lower limit of the time.
Return
Format
The query returns GRE, LESS, or GLES.
Example
:TRIGger:DURATion:WHEN LESS /*Set the trigger condition to <*/
:TRIGger:DURATion:WHEN? /*The query returns LESS*/
:TRIGger:DURATion:TUPPer
Syntax
:TRIGger:DURATion:TUPPer <NR3>
:TRIGger:DURATion:TUPPer?
Description
Set or query the duration time upper limit in duration trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<NR3>
Real
Related to the trigger condition
LESS: 8ns to 10s
GLESs: 16ns to 10s
2μs
Explanation
This command is available when the trigger condition (:TRIGger:DURATion:WHEN) is
LESS or GLESs.
Return
Format
The query returns the duration time upper limit in scientific notation.
Example
:TRIGger:DURATion:TUPPer 0.000003 /*Set the duration time upper limit to 3μs*/
:TRIGger:DURATion:TUPPer? /*The query returns 3.000000e-06*/
Related
Command
:TRIGger:DURATion:TLOWer

Table of Contents

Other manuals for Rigol DS1104Z-S Plus

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z-S Plus and is the answer not in the manual?

Rigol DS1104Z-S Plus Specifications

General IconGeneral
Bandwidth100 MHz
Sample Rate1 GSa/s
Channels4
Memory Depth24 Mpts
Vertical Resolution8 bits
Vertical Sensitivity1 mV/div to 10 V/div
Timebase Range2 ns/div to 1000 s/div
Waveform RecordUp to 60, 000 frames
FFTYes
Waveform Capture Rate30, 000 wfms/s
Display7-inch TFT LCD
Trigger ModesEdge, Pulse, Video, Window, Timeout
InterfacesUSB, LAN
Serial Bus Triggering and DecodingI2C, SPI, UART
Input Impedance1 MΩ
Power Supply100-240 VAC, 45-440 Hz