EasyManuals Logo
Home>Rigol>Test Equipment>DS1104Z-S Plus

Rigol DS1104Z-S Plus Programming Guide

Rigol DS1104Z-S Plus
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #198 background imageLoading...
Page #198 background image
RIGOL Chapter 2 Command System
2-182 MSO1000Z/DS1000Z Programming Guide
:TRIGger:DURATion:TLOWer
Syntax
:TRIGger:DURATion:TLOWer <NR3>
:TRIGger:DURATion:TLOWer?
Description
Set or query the duration time lower limit in duration trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<NR3>
Real
8ns to 9.99s
1μs
Explanation
This command is available when the trigger condition (:TRIGger:DURATion:WHEN) is
GREater or GLESs.
Return
Format
The query returns the duration time lower limit in scientific notation.
Example
:TRIGger:DURATion:TLOWer 0.000003 /*Set the duration time lower limit to 3μs*/
:TRIGger:DURATion:TLOWer? /*The query returns 3.000000e-06*/
Related
Command
:TRIGger:DURATion:TUPPer

Table of Contents

Other manuals for Rigol DS1104Z-S Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z-S Plus and is the answer not in the manual?

Rigol DS1104Z-S Plus Specifications

General IconGeneral
BrandRigol
ModelDS1104Z-S Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals