RIGOL Chapter 2 Command System
2-188 MSO1000Z/DS1000Z Programming Guide
:TRIGger:RUNT:BLEVel
Syntax
:TRIGger:RUNT:BLEVel <level>
:TRIGger:RUNT:BLEVel?
Description
Set or query the trigger level lower limit in runt trigger. The unit is the same as the current
amplitude unit.
Parameter
Name
Type
Range
Default
<level>
Real
(-5 x VerticalScale - OFFSet) to
(5 x VerticalScale - OFFSet)
0
Explanation
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return
Format
The query returns the trigger level lower limit in scientific notation.
Example
:TRIGger:RUNT:BLEVel 0.16 /*Set the trigger level lower limit to 160mV*/
:TRIGger:RUNT:BLEVel? /*The query returns 1.600000e-01*/