4 To Set the Sample System RIGOL
DS6000 User’s Guide 4-5
Sample Mode
Press ACQ  Sampling and press this softkey continuously to switch the sample
mode (the default is real-time sample). Note: All the "Acquisition Mode" support
real-time sample and equivalent sample.
Real-time Sample
In this mode, the oscilloscope samples and displays waveform within a trigger event.
The maximum real-time sample rate of DS6000 is 5 GSa/s and the current sample
rate is displayed under the Sa Rate menu.
Equivalent Sample
In this mode, the oscilloscope performs multiple trigger samples, samples at
different positions of each trigger and uses the sample points from multiple triggers
to rebuild the signal. The equivalent sample mode helps with detailed observation of
the repeated periodic signals. The maximum equivalent sample rate of DS6000 is
100 GSa/s.
Tip
Press Run/Stop to stop the sample, the oscilloscope will hold the last display. At
this point, you can still use the vertical control and horizontal control to pan and
zoom the waveform.

Table of Contents

Other manuals for Rigol DS6062-DEMO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS6062-DEMO and is the answer not in the manual?

Rigol DS6062-DEMO Specifications

General IconGeneral
BrandRigol
ModelDS6062-DEMO
CategoryTest Equipment
LanguageEnglish

Related product manuals