3 Basic Operation
3.5
3.
18 FPD-2010
3.5
Checking Glass Insert and Injection Port Septum
This section describes how to check the glass insert. For the detailed procedure, refer to "12. Injection port"
in the GC-2010 Instruction Manual.
Confirm that the glass insert suits the purpose.
 WARNING
High temperature
You might suffer burns. Be sure to confirm that the sample injection port has
cooled down before staring the operations.
For split For splitless
Fig. 3.5.1 
Replace the injection port septum around when the times of injection reach 100.