EasyManuals Logo
Home>Siemens>Measuring Instruments>SD 270

Siemens SD 270 User Manual

Siemens SD 270
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
36
5 Specification
5.1 Output data
Outputs: Left earphone, Right earphone,
Bone (L&R) Insert masking and Freefield
Frequency range (Hz): Air: 125-8kHz
Bone: 250Hz-8kHz
Frequency accuracy: <1%
Distortion: <2%
Output level range (AC): -10dBHL to 120dBHL maximum
Output level range (BC): -10dBHL to 70dBHL maximum
Output level range (FF): Up to 90dB
Insert masking output: 90dBHL max (250-4kHz)
Output level accuracy: Within 3dB
Output level step size: 2.5 or 5dB
Output transducer (AC): DD45 earphones (supplied)
E-5A insert earphones (option)
Output transducer (BC): B-71 bone vibrator (supplied)
Tone present: Single, pulsed, warble or continuous
Masking: Narrowband or broadband
Lock: Synchronous channel lock
Clinical tests: SISI, Stenger ABLB
Communication: Integral talk over and talk back facility
Recorded speech: Tape or CD input
Live speech: 1 x microphone input
Monitoring indicator: VU ‒ (to IEC 60268-17)
NOAH3 interface: Transfer of AC, BC, ULL and speech
test results