HRX-OM-K003
Chapter 4 System Startup and Shutdown
HRW Series 4.5 System Startup and Shutdown
4.4.2 Circulating fluid temperature setting
From the “Setting screen” on the LCD screen, set the circulating fluid at any
temperature. See section 5.4 “Examples of System Operation” in “Chapter 5
System Operation” on page 5-11 for operating procedure.
[Tips]
See section 8.1.1 “System specification” in “Chapter 8 Appendix” on page 8-1
for the setting range of circulating fluid temperature.
4.5 System Startup and Shutdown
4.5.1 System startup
Press the [START/STOP] key on the operation display panel.
The ‘RUN’ lamp on the operation display panel comes on, and system operation is initiated
4.5.2 System shutdown
Press the [START/STOP] key on the operation display panel.
‘RUN’ lamp on the operation display panel goes out, and the system comes to a halt accordingly.
Internal equipment may remain at elevated in temperature immediately
after system shutdown. Potential burns may happen if your skin
comes in contact with these surfaces. Further work is allowed only
when the system reaches room temperature.
Emergency off [EMO] switch and main breaker (OFF) should not be
used for system shutdown unless it is an emergency.