A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
1-6
1.8 Electrical Safety Classification
¾ This safety standard of this system complies with Type B, Class I
EN606601-1:1990+A1:1993+A2:1995;
¾ It belongs to continuous operation (for more than 8 hours)
according to operation mode;
¾ According to the classification on harmful liquid
protection(Main unit:IPX0 Transducers:IPX4)
¾ The system should NOT be used under the condition with the
inflammable mixture of anesthesia gas and air, or oxygen and
nitrous oxide;
¾ This equipment meets the requirements for EMC/EMI
IEC60601-1-2 Class A (CE);
¾ The system shall be added with a ground device protection.
During operation, please make sure that the system is connected
to an extra grounding device. Otherwise it may be interfered by
alternating current and affected in imaging quality.