A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
5-2
5.1.2 Installation
The system is a portable diagnostic device.
Before installation, please pay attention to the following:
z The environment where the system is in use shall comply with the
requirements in its technical specifications;
z Take out the system from the package; ensure that the system
cosmetic is good, without any obvious damage;
z The system shall be placed on a even platform, without any irrelevant
items around;
z The grounding post at the back of the system shall be connected to a
terminal complying with grounding standard.
5.1.3 System Transport
When moving or transporting the system, take the precautions outlined below
to ensure maximum safety for personnel, the system, and other equipment.
After the system is positioned to a new place, ensure that the system is
placed on a secure level.
Before moving or transporting the system, observe the following:
1. Completely power down the system; plug out the power cord.
2. Disconnect all the transducers, and store all transducers in their
original cases;
3. Disconnect all peripheral devices;
4. Lift the system when all the above items are observed.
5. For long-distance transport, observe the additional precautions below:
6. Pack the system properly. It is recommended to pack it with its original
package, so as to avoid impact, shock or pressure during transport;
7. Do not place the system upside down during transport;
8. For environmental requirements on transport, see relevant sections in
this manual.