A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
5-7
To ensure efficacy of the disinfectant, clean the transducer thoroughly before
disinfection. Make sure that not any residue is on the transducer. Here are the
disinfection steps for a transducer:
z According to the precautions for storage, usage and handling of the
disinfectant, follow the supplier’s instruction to prepare the
disinfectant.
z Immerse the cleaned and dried transducer in the prepared
disinfectant following the requirement.
z When the transducer is immersed or placed into the disinfectant,
rotate or shake the transducer section in the disinfectant to clear off
any bubble and ensure a thorough contact between the transducer
and the disinfectant. For high-level disinfection, perform the
disinfection procedure by following the supplier’s suggested duration.
z Withdraw the transducer from the disinfectant, and the disinfection for
the transducer is finished. Clean off the disinfectant on the transducer
thoroughly with water, and then wipe it up with soft dry cloth.
z Air dries the transducer.
5.3 Service Method
5.3.1 Service Method
When any abnormality occurs, shut off the power supply and inform our
service engineer of the detailed situation. If possible, send a failure picture to
our service engineer. We will provide prompt service. A detailed failure report
will help us to shorten the service time.
If the fuse is broken, the fuse can be replaced by following our service
engineer’s instruction. The fuse is 50T-0160H 250V. Two fuses of this model
were packed with the system as accessories. The same fuses can be
purchased from the local market. If the same broken fuse issue happens
again after replacement, please inform our service engineer immediately.