A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-39
B3-Axis
Function: Based on the vertical section and the horizontal section of the
measured object, the object volume is calculated.
Principle: The volume can be estimated from the orthogonal D1, D2 and D3
values of the object.
The formula:
V = (π/6) × D1 × D2
× D3
4. Move the trackball to adjust the variable axis
length of the ellipse to match the ellipse with
the measured region. Move the trackball
leftward, and the variable ellipse axis is
getting small, Move the trackball rightward,
and the variable ellipse axis is getting bigger.
5. Press theSETkey to confirm the ellipse
measurement area, and the result is
displayed in the result window, and it ends
the measurement.
6. Repeat steps 2 to 5 to perform the next
measurement.