A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-75
z Simpson (Simpson)
Measurement items: LVAMd, LVLd, LVAPD, LVAMs, LVLs, LVAPs;
Calculation items: EDV, ESV, SV, CO, EF, SI, CI
Use the formula:
EDV=
(
)
LVAPdLVAMdLVAPdLVAMd
LVLd
×+×+×× 24
9
ESV=
(
)
LVAPsLVAMsLVAPsLVAMs
LVLs
×+×+×× 24
9
Measurement method:
1 In B mode freeze the image, press the Calc
key to pop up the Meas & Calc menu, move the
trackball, use the cursor to select “Cardiac”,
press theSETkey to go to the submenu,
move the trackball to select “LV” , press the
SETkey to go to the submenu, move the
trackball to select “Simpson”, press theSET
key to go to measurement state.
2 Measure LVAMd, see “Area/Cir measurement
(Ellipse)” for the specific measurement method.
3 Measure LVLd, see “Distance Measurement”
for the specific method.
4 Measure LVAPD; see “Area/Cir measurement
(Ellipse)” for the specific measurement method.
5 Unfreeze the image, rescan the image, and
freeze the image.
6 Measure LVAMs, see “Area/Cir measurement
(Ellipse)” for the specific measurement method.
7 Measure LVLs, see “Distance Measurement”
for the specific method.