General
lnformation-2215A
Operators
Table 1-1 (cont)
Characteristics
Performance Requirements
Supplemental Information
HORIZONTAL DEFLECTION SYSTEM
Sweep Rate
Calibrated Range
A Sweep
0.5 s per division
to
0.05
µs
per
division in a 1-2-5 sequence. X10
magnifier extends maximum sweep
speed
to
5
ns
per division.
B Sweep
50
ms
per division to 0.05
µs
per
division
in
a 1-2-5 sequence. X10
magnifier extends maximum sweep
speed
to
5
ns
per division.
Accuracy
Unmagnified
Magnified
Sweep accuracy applies over the
center 8 divisions. Exclude the first
+ 15°C
to
+ 35°C
± 3%
± 4%
25
ns
of the sweep for magnified
sweep speeds and anything beyond
0°c
to
+ 50°c
± 4%
± 5%
the 100th magnified division.
POSITION Control Range
Start
of
sweep
to
10th division in X 1
or
100th divisions in X10 will position past
the center vertical graticule line.
Sweep Linearity
± 7%.
Linearity measured over any 2
of
the
center 8 divisions. With magnifier in X10,
exclude the first 25
ns
and anything past
the 100th division.
Variable Control Range
Continuously variable between
calibrated settings. Extends the A and
B sweep speeds by
at
least a factor
of
2.5.
Delay Time
Applies to 0.5
þÿ¼s
per division
Delay time
is
functional
but
not
and slower.
calibrated
at
sweep settings above 0.5
þÿ¼s
per division.
Dial Control Range
< 0.5 + 300 ns
to
> 10 divisions.
Jitter
One part
or
less
in
20,000 (0.005%)
of
the maximum available delay time.
Differential Time Measurement
Exclude delayed operation when A and B
Accuracy
SEC/DIV knobs are locked together
at
+ 15°C
to
+ 35°c
± 1 %
+0.015
major dial division
any sweep speed
or
when A SEC/DIV
switch is
at
0.5
þÿ¼s
per division
or
faster.
o· c
to
+ 50°c
±
2%
+0.015
major dial division.
Acc
ur
acy applies over the B DELAY TIME
POSITION control range.
1-4

Table of Contents

Other manuals for Tektronix 2215A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2215A and is the answer not in the manual?

Tektronix 2215A Specifications

General IconGeneral
BrandTektronix
Model2215A
CategoryTest Equipment
LanguageEnglish

Related product manuals