Installation
Powering the Instrument On and Off
This instrument operates from a single-phase power source with the neutral conductor at o r near earth ground. (See
page 1, Operating C onsiderations.) A protective ground connection through the grounding conductor in the power cord is
essential for safe operation.
Functional Check
1. Power on the instrument
2. Push User Opt > Test.
3. Push the Test Mode menu button and select All.
4. Push the Start test menu button. Verify that all tests pass.
5. Push the OK menu button to return to the previous measurements display.
6. Connect a 50 Ω cable from the 10 MHz Output connector on the rear panel to the Input A connector on the front panel.
7. Push Input A and select 50 Ω input impedance.
8. Push Meas > Freq > Freq > A . The instrument should read 10 MHz.
2 FCA3000, FCA3100, and MCA3000 Series Quick Start User Manual