Arguments
<NR3> is the A trigger holdoff time. The range is 20 ns to 8.0 s.
Examples
TRIGger:A:HOLDOff:TIMe10 sets the holdoff time to 10 s.
TRIGGER:A:HOLDOFF:TIME? might return :TRIGGER:A:HOLDOFF:TIME
1.2000E-06 indicating that the A trigger holdoff time is set to 1.2 μs.
TRIGger:A:LEVel
Sets or queries the trigger level for the A trigger. This command is equivalent to
adjusting the front-panel TRIGGER LEVEL knob.
NOTE. When the edge trigger source is set to AC LINE, the instrument ignores
the set form of the command.
When the edge trigger source is set to AC LINE, the query form of the command
returns zero.
Group
Trigger
Syntax
TRIGger:A:LEVel{ECL|TTL|<NR3>}
TRIGger:A:LEVel?
Arguments
<NR3> specifies the trigger level in user units (usually volts).
ECL specifies a preset ECL high level of -1.3V.
TTL specifies a preset TTL high level of 1.4V.
Examples
TRIGGER:A:LEVEL TTL sets the A edge trigger to TTL high level, which is
1.4 V.
TRIGger:A:LEVel? might return :TRIGGER:A:LEVel 1.3000E+00 indicating
that the A edge trigger is set to 1.3 V.
T commands
228 TBS2000 Series Programmer

Table of Contents

Other manuals for Tektronix TBS2000 Series

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
Bandwidth70 MHz, 100 MHz, 200 MHz
Channels2 or 4
Time Base Range5 ns/div to 50 s/div
Probe InterfaceTekVPI
Operating Temperature0 °C to +50 °C
Vertical Resolution8 bits
Sample Rate1 GS/s
Memory Depth20 kpts
Record Length20 kpts
Display9-inch WVGA display
Input Impedance1 MΩ ±1%, 13 pF ±2 pF
Vertical Sensitivity2 mV/div to 5 V/div
Trigger TypesEdge, Pulse Width, Video
ConnectivityUSB, LAN
Math FunctionsAddition, Subtraction, Multiplication, Division, FFT

Summary

Getting Started with TBS2000 Series Programmer

Remote Operation Overview

Provides information on remotely controlling the instrument via communication protocols and commands.

Understanding Command Syntax

Command and Query Structure

Explains the formation of set and query commands, including headers, mnemonics, and arguments.

BNF Notation

Details the Backus-Naur Form used for describing command syntax and structure.

Command Entry Rules

Outlines general rules for entering commands, including case sensitivity and whitespace.

Concatenating Commands

Explains how to combine multiple commands and queries using semicolons for sequential execution.

Organizing Commands into Groups

Acquisition Command Group

Manages waveform acquisition modes, settings, and averaging parameters.

Ethernet Command Group

Controls the instrument's Ethernet interface settings and operations.

Horizontal Command Group

Manages horizontal axis settings, time base, and record length.

Measurement Command Group

Configures and retrieves automated measurement results.

Trigger Command Group

Defines and manages instrument trigger conditions and settings.

Understanding Status and Events

Registers Overview

Explains the instrument's status and enable registers for event reporting.

Event Handling Sequence

Details the process of how the status and event handling system manages instrument events.

Related product manuals