EasyManuals Logo
Home>Tektronix>Test Equipment>TBS2000 Series

Tektronix TBS2000 Series User Manual

Tektronix TBS2000 Series
328 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #270 background imageLoading...
Page #270 background image
WFMInpre:XINcr
Sets or queries the horizontal interval between incoming waveform points in
units specified by WFMInpre:XUNit.
Group
Waveform
Syntax
WFMInpre:XINcr <NR3>
WFMInpre:XINcr?
Arguments
<NR3> is the interval between points in the waveform record, in the units
specified by WFMPre:XUNit.
Examples
WFMINPRE:XINCR 3E-3 sets the interval between incoming waveform points
to 3 ms.
WFMINPRE:XINCR ? might return :WFMINPRE:XINCR 1.0000E-3 indicating
that if WFMInpre:XUNit is set to "s", there is a 1 ms interval between incoming
waveform points.
WFMInpre:XUNit
Sets or queries the horizontal units of the incoming waveform.
Group
Waveform
Syntax
WFMPre:XUNit <Qstring>
WFMPre:XUNit?
W commands