CLEARMenu
Clears the current menu from the display. This command is equivalent to
pressing the front panel Menu off. No query form.
Group
Miscellaneous
Syntax
CLEARMenu
Examples
CLEARMenu clears the current menu from the display.
*CLS
Command only, no query form. The *CLS command clears the following
instrument status data structures:
â– 
The Event Queue
â– 
The Standard Event Status Register (SESR)
â– 
The Status Byte Register (except the MAV bit)
If the *CLS command immediately follows an <EOI>, the Output Queue and
MAV bit (Status Byte Register bit 4) are also cleared. MAV indicates
information is in the output queue. The device clear (DCL) GPIB control
message and the USBTMC INITIATE_CLEAR control message will clear the
output queue and also MAV.
*CLS does not clear the output queue or MAV. *CLS can suppress a service
request that is to be generated by an *OPC command. This will happen if a hard
copy output or single sequence acquisition operation is still being processed
when the *CLS command is executed. See Registers on page 275
Group
Status and Error
C commands
TBS2000 Series Programmer 75

Table of Contents

Other manuals for Tektronix TBS2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS2000 Series and is the answer not in the manual?

Tektronix TBS2000 Series Specifications

General IconGeneral
BrandTektronix
ModelTBS2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals