Programable DC Electronic Load User Manual
19
4.1 Constant State Test Modes
There are four types of constant state test modes: constant current (CC), constant voltage
(CV), constant resistance (CR) and constant power (CP). In the initialization interface,
users can select [Mode] by pressing the soft key at the bottom of the screen, and then
press [CC], [CV], [CR], [CP] to enter the corresponding mode.
Users can also press [More] to select other modes, see Table 4-1.
Table 4-1 Introduction of Constant State Test Modes
Name Description
CC mode
The electronic load always consumes constant current regardless of the
input voltage.
CV mode
The electronic load changes current consumption to keep the input
voltage as the set value.
CR mode
The electronic load is equivalent to a resistor.
current as the voltage changes.
CP mode
The electronic load always consumes constant power. It reduces the
current when the voltage rises to remain constant power.
More modes
There are many modes under this menu that can be selecte
cursor to meet diverse testing requirements.
V, A, W
parameters
Present voltage, current and power value
Status bar Current mode, time, status and other parameters
In constant state test modes, set current/voltage/resistance/power and other parameters
will be displayed on the right side of screen. Input on-load parameters that need to be set
by numeric keyboard or knob. In on-load status, Time, Vpp, Ipp and other parameters will
be displayed at the bottom of the main screen.
4.1.1 Constant Current Test
In CC mode, the electronic load always consumes constant current regardless of the input
voltage.