Programable DC Electronic Load User Manual
24
OVP, Short Circuit, Time and more, as shown in Table 4-2.
Table 4-2 Introduction of More Modes
More modes Description
Dynamic
Two different voltage/current values are set. The load will switch between
these two values when testing.
List
Up to 16-step different loading modes can be set. It can customize the
stepping mode and the upper and lower limits of test and judgment, and
have the function of saving and memorizing for each group of test modes.
Dual
Two constant test modes combine to form a dual on-load mode, including
CR+CC, CV+CR and CV+CC.
OCP/OPP
Set the overload current/power/voltage value to identify whether the
protection function of the tested object is normal. The load continuously
increases the set load value during the test, and detects the value of the
tested object when performing the protection function.
CR-LED
The test mode is for LED power supply, which can simulate the operating
parameters (voltage/current and LED coefficient) to measure the
performance of LED power supply.
Battery
Discharge the measured battery by CC/CR/CP mode. The test will end
automatically when reaching the ending value. The battery capacity and the
discharge graph will also be display.
Load effect
Load according to three different loads of the set value to calculate ΔV and
Reg (load regulation).
OVP
Set the overload current/power/voltage value to identify whether the
protection function of the tested object is normal. The load continuously
increases the set load value during the test, and detects the value of the
tested object when performing the protection function.
Short circuit
A short-circuited circuit will be simulated in input terminal to test if the
protection function of measured object can work normally when its output
terminal is short-circuited.
Time
For the start/drop time test of the switching power supply, set the start value
and end value of the load. After the voltage of the tested object is stabilized,
the time it takes for the process to rise to the stable voltage is detected.
4.2.1 Dynamic Mode
UTL8500X series electronic load has two dynamic loading modes: CC and CV. In the
dynamic mode, users can set two fixed-value parameters for corresponding constant state