Programable DC Electronic Load User Manual
31
Figure 4-2-3-1 Interface of Dual Mode
Parameter setting in dual mode:
Parameter
Description
Mode
CR+CC/CV+CR/CV+CC
Start value Set the start value of the dual mode
Swap value Set the swap value of the dual mode
To use the dual mode, first select the needed mode, and then set the fixed values of the
two corresponding modes. When the load starts to load, first start the load according to
the fixed value of the first mode, and when the external input changes so that the
parameter reaches the set switching value, the load switches to the second load mode.
Figure 4-2-3-2 Dual Mode
Description of dual mode:
When using dual mode, it is necessary to set reasonable starting and switching
parameters to ensure that the set switching value can be reached during the test.
Operation steps:
1. In initial interface, select [Mode], [More], and then press [Dual] to enter the setting
interface.
2. Move the cursor to the needed setting position by pulse knob, and press the Enter
Constant
resistance
Constant current

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the UNI-T UTL8500X Series and is the answer not in the manual?

UNI-T UTL8500X Series Specifications

General IconGeneral
BrandUNI-T
ModelUTL8500X Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals