14. Technical Specifications
Green Smart User Manual 139
ENGLISH
14.2 Technical Specifications
14.2.1 X-ray Generator Specifications
Specifications
Item Description
Generator
Model
DG-07E22T2
Rated output power
1.6kW
Inverter model
name
INV-22
Type
Inverter
Normal/
Pulse
kVp
60 kV ~ 99 kV
(1 kV increment)
mA
4 mA ~ 16 mA
(0.1 mA increment for CBCT,
1 mA increment for PANO and CEPH)
Cooling
Cooling Protect
(Optional fan cooling ≥ 60
)
Total filtration
Min. 2.5 mm Al
Default filtration
1.0 mm Al
Added filtration
1.5 mm Al (Fixed) / PANO and CEPH
mode
1.5 mm Al (Fixed) + 3.0 mm Al
(Automatically added) / CBCT mode
Tube
Manufacturer
Toshiba
Model
D-052SB (Stationary Anode type)
Focal spot size
0.5 mm (IEC 60336)
Target Angle
5 degree
Inherent Filtration
At least 0.8 mm Al equivalent at 50 kV
X-ray Coverage
95 mm x380 mm at SID 550 mm
Anode Heat
Content
35 kJ
Duty Cycle
1:60 or more
(Exposure time: Interval time)

Table of Contents