EasyManuals Logo
Home>Vatech>Medical Equipment>GREEN SMART

Vatech GREEN SMART User Manual

Vatech GREEN SMART
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #88 background imageLoading...
Page #88 background image
7. Acquiring PANO Images
80 Green Smart User Manual
7.4 X-ray Exposure
If an emergency occurs during image acquisition, release the
Exposure Switch to cease X-ray emission.
The operator shall observe the X-ray safety regulations applicable
to his/her area at all times during the operation of this equipment.
The operator must keep vocal/visual contact with the patient at all
times during the image acquisition process.
Do not operate the PC during exposure. Doing so may cause the
system to malfunction.
Let the patient close the eyes during the operation.
To acquire optimized images, instruct the patient to hold his/her
breath and not to swallow. Also, don’t let the patient move until the
Temple Supports are open.
1. Get out of the X-ray room and close the door.
The operator must keep vocal/visual contact with the patient at all
times during image acquisition.
2. Press and hold down the Exposure Switch until image acquisition is completed.
The image appears on the screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech GREEN SMART and is the answer not in the manual?

Vatech GREEN SMART Specifications

General IconGeneral
BrandVatech
ModelGREEN SMART
CategoryMedical Equipment
LanguageEnglish

Related product manuals