3. Warnings and Precautions
Green Smart User Manual 15
ENGLISH
3.3 Radiation Safety
Since rules and regulations concerning radiation safety differ between
countries, it is the responsibility of the owner and/or operator of this
equipment to comply with all applicable rules and regulations
concerning radiation safety and protection in his/her area.
This equipment must be housed inside an X-ray shielded room.
The operator must remain outside a shielded room during X-ray exposure to protect
himself/herself from radiation.
During imaging; the operator must maintain vocal/visual contact with the patient from
outside the shielded area.
The operator should continuously check the status of the patient and the equipment
during imaging.
The operator should be at least 2 m (6 feet) away from the equipment during imaging.
The operator must immediately stop imaging if the equipment malfunctions.
The patient must wear a lead apron with neck and thyroid protection during X-ray
exposure.
Children and pregnant women must consult with a doctor before X-ray exposure.
As a manufacturer of radiology equipment that conforms to stringent
protection standards around the world, we guarantee the maximum
degree of protection against radiation hazards for our equipment.