EasyManuals Logo
Home>Xylem>Measuring Instruments>wtw Multi 3410

Xylem wtw Multi 3410 User Manual

Xylem wtw Multi 3410
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
Operation Multi 3410
37
ba75863y18 01/2015
4.6 pH value
4.6.1 General information
You can measure the following parameters:
pH value [ ]
Sensor voltage [mV]
Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This enables interference-free measurements in
grounded measuring media.
Temperature
measurement
For reproducible pH measurements, it is essential to measure the tem-
perature of the test sample. IDS sensors measure the temperature with
a temperature sensor integrated in the IDS sensor.
Preparatory activities Perform the following preparatory activities when you want to measure:
4.6.2 Measuring the pH value
When operating a sensor without integrated temperature sensor, e.g.
via an IDS pH adapter, you have to determine and input the tempera-
ture of the measuring medium first.
1 Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.
2 If necessary, select the measured parameter with <M>.
3 If necessary, calibrate or check the IDS pH sensor.
1 Perform the preparatory activities according to section 4.6.1.
2 Immerse the IDS pH sensor in the test sample.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem wtw Multi 3410 and is the answer not in the manual?

Xylem wtw Multi 3410 Specifications

General IconGeneral
BrandXylem
Modelwtw Multi 3410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals