Operation Multi 3410
65
ba75863y18 01/2015
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.8.3 Settings for D.O. sensors
(menu or measurement and calibration settings)
Overview The following settings are possible for D.O. sensors:
Calibration record (display)
Calibration data storage (display)
Calibration interval
FDO Check
Check interval
Salinity correction
Salinity (salinity equivalent)
Settings The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the <MENU/ENTER> key. After
completing the settings, switch to the measured value display with
<M>.
Measured
parameter
Time interval Stability in the time interval
D.O. concen-
tration
20 seconds
Δ
: Better than 0.03 mg/l
D.O. saturation 20 seconds
Δ
: Better than 0.4 %
D.O. partial
pressure
20 seconds
Δ
: Better than 0.8 mbar
Temperature 15 seconds
Δ
: Better than 0.5 °C
Menu item Possible
setting
Description
Calibration /Cali-
bration record
- Displays the calibration record of the
last calibration
Calibration /Cali-
bration data
storage
- Displays the last calibration records
(max. 10)