Multi 3410 Operation
56
ba75863y18 01/2015
4.6.10 QSC function (sensor quality control)
General information
on the QSC function
The QSC function (Quality Sensor Control) is a new sensor evaluation
for digital IDS sensors. It evaluates the condition of an IDS pH sensor
individually and with a very fine grading.
On the display, the QSC color scale (from green to yellow) indicates the
current sensor evaluation by means of a pointer.
In the printout, the sensor evaluation is quoted as a percentage
(1-100).
The finely graded sensor evaluation of the QSC function promptly calls
your attention to changes of the sensor.
Thus you can do what is necessary to restore the optimum measuring
quality (e.g. clean, calibrate or replace the sensor).
Sensor evaluation
with / without
QSC function
With QSC function Without QSC function (sensor
symbol)
Very fine grading of the sensor
evaluation (100 grades)
Rough grading of the sensor
evaluation (4 grades)
The reference value is individu-
ally determined for each sensor
during the QSC initial calibra-
tion.
A theoretical reference value is
used for all sensors
Low tolerances for zero point
and slope when using QSC buf-
fer solutions
Greater tolerances for zero point
and slope when using commer-
cial buffer sets
Additional QSC calibration
required (with special QSC buf-
fer set)
No additional calibration required

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem wtw Multi 3410 and is the answer not in the manual?

Xylem wtw Multi 3410 Specifications

General IconGeneral
BrandXylem
Modelwtw Multi 3410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals