Operation Multi 3410
41
ba75863y18 01/2015
4.6.4 pH calibration
Why calibrate? During the operation of an IDS pH sensor, the zero point (asymmetry)
and slope of the sensor change with time.As a result, an inexact mea-
sured value is displayed. Calibration determines the current values of
the zero point and slope of the pH sensor and stores them in the mea-
suring instrument.
Thus, you should calibrate at regular intervals.
When do you have to
calibrate?
When the calibration interval has expired
Buffer sets for
calibration
You can use the buffer sets quoted in the table for an automatic calibra-
tion. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into consideration
during the calibration.
mV resolution 0.1
1
Resolution of the mV display
Reset - Resets all sensor settings to the delivery
condition (see section 4.14.1)
Menu item Possible setting Description
No. Buffer set* pH values at
1
VariCal Any Any
2
NIST/DIN
DIN buffers according to DIN
19266 and NIST Traceable Buf-
fers
1.679
4.006
6.865
9.180
12.454
25 °C
3
TEC
WTW Technical buffers
2.000
4.010
7.000
10.011
25 °C
4
Merck 1* 4.000
7.000
9.000
20 °C
5
Merck 2 * 1.000
6.000
8.000
13.000
20 °C
6
Merck 3 * 4.660
6.880
9.220
20 °C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem wtw Multi 3410 and is the answer not in the manual?

Xylem wtw Multi 3410 Specifications

General IconGeneral
BrandXylem
Modelwtw Multi 3410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals