EasyManuals Logo
Home>YOKOGAWA>Measuring Instruments>ADMAG AXR

YOKOGAWA ADMAG AXR User Manual

YOKOGAWA ADMAG AXR
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #107 background imageLoading...
Page #107 background image
<8. OPERATION VIA HART CONFIGURATION TOOL (HART 5)>
8-2
IM 01E30D01-01EN
(2) ConrmtheinstalledDDrevisionin
accordance with the procedure of the
CongurationTool.Readitsmanualhowto
conrmitindetail.
8.3 InterconnectionbetweenAXRandHARTCongurationTool
TheHARTCongurationToolcaninterfacewiththeAXRfromthecontrolroom,theAXRsite,oranyother
wiringterminationpointintheloop,providedthereisaminimumloadresistanceof250Ωbetweenthe
connectionandthereceivinginstrument.Tocommunicate,itmustbeconnectedinparallelwiththeAXR,and
the connections must be non-polarized. Figure 8.3 illustrates the wiring connections for a direct interface at the
AXRsite.TheHARTCongurationToolcanbeusedforremoteaccessfromanyterminalstripaswell.
+
SUPPLY
DO
USB
FieldMate
modem
USB
PC/FieldMatePC/FieldMate
AXR
F0802.ai
PC/FieldMate
4 to 20 mA DC signal line
Relaying
terminals
Terminal board
Control room
Receiving
instrument
load resistance:
250 to 600
Distributor
Figure 8.3 Connecting the HART Communicator
WARNING
Be sure to set parameters as “Protect” on the
writeprotectfunctionafternishofparameter
setting work.
Read Chapter 6 Menu P and Subsection 11.3.2,
how to use the write protect function in detail.
IMPORTANT
Ifthepowerofowmeteristurnedoffwithin30
seconds after parameters have been set, these
settings will be canceled. Accordingly, please
keep the power on for at least 30 seconds after
setting parameters.
NOTE
Before updating any setting, remember to always
check the parameter content you want to change
as described in Chapter 6.
NOTE
IncaseofHARTConguration,theparameters
are displayed in English only.
Even if the language with the exception of
English is selected at “Language” setting display,
the parameters are displayed in English upon
HARTCongurationTool.
8.2.2 Device Type Manager (DTM) and
Device Revision
WhenconguretheparametersusingFieldMate,
use the DTM (Device Type Manager) reading the
following table.
MODEL DTM
Name
Device
Type
Device
Revision
AXR AXRV2.1 AXR(0x57) 2
The device revision can be checked as follows.
DTM Conguration→HART→Flddev
rev

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXR and is the answer not in the manual?

YOKOGAWA ADMAG AXR Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals