<5. BASIC OPERATING PROCEDURES (USING THE DISPLAY UNIT)>
5-8
IM 01E30D01-01EN
F0505-2.ai
Major item
Sub-item
selection (B)
Incorrect setting:
Incorrect setting:
Out of range.
Touch any key.
Invalid value.
Touch any key.
Sub-item
Parameter
Search Mode
B23:Flow Span
120 l/min
B:Easy Setup
22:Base Time Unit
23:Flow Span
24:Flow Decimal Pnt
B23:Flow Span
100 l/min
1200000 l/min
Rng:0.00001 32000
B23:Flow Span
100 l/min
0001.0. l/min
Rng:0.00001
32000
SET
When the
SET
switch is touched, the entire
displayashesonandoff.Conrmthatthesetting
has been correctly changed to “120,” and then
xthisvaluebytouchingthe
SET
switch once
again.
NOTE
When no operations are carried out for 20
secondsintheashingstate,thesystemwill
automatically return to the Sub-item Parameter
Search Mode.
When the operations except
SET
are carried
out, the parameter cannot be set.
The system automatically returns to sub-item
selection screen (B).
NOTE
If the input value is outside the valid selection
range, the message “Out of range. Touch any
key.” will be displayed. In such a case, touch
any switch to return to Parameter Replacement
Mode and redo the setting.
NOTE
If more than one decimal point has been input,
the message “Invalid value. Touch any key.” will
be displayed. In such a case, touch any switch
to return to Parameter Replacement Mode and
redo the setting.
5.3.3 Setting Example for Alphanumeric-Type Data: Tag number
This example describes the setting of the tag number for the alphanumeric-type parameter C10: Tag No. from
“FI-1101” to “FI-1201.”
Setting mode
P:Protect
B:Easy Setup
C:Basic Setup
Setting mode
B:Easy Setup
C:Basic Setup
D:Total Set
INC
SET
F0506-1.ai
Setting mode
Major item
parameter
Setting mode
Major item
parameter
Start:
Major Item
Parameter
Search Mode
Setting Mode Condition
Touch the
INC
switch to access C: Basic
Setup.
NOTE
The▲and▼symbolstotheleftofthe
parameters indicate that additional setting items
to those being currently displayed may also be
selected. Use the
INC
switch to cycle through
these items.
The cursor has been moved to C: Basic Setup in
this screen.
Touch the
SET
switch to enter C: Basic Setup.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXR and is the answer not in the manual?

YOKOGAWA ADMAG AXR Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals