EasyManuals Logo
Home>YOKOGAWA>Measuring Instruments>ADMAG AXR

YOKOGAWA ADMAG AXR User Manual

YOKOGAWA ADMAG AXR
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
<5. BASIC OPERATING PROCEDURES (USING THE DISPLAY UNIT)>
5-9
IM 01E30D01-01EN
Selection of the appropriate parameter
C:Basic Setup
49:Flow User Span
10:Tag No
11:Flow Damping
C10:Tag No
Fl-1101
l-1101
C10:Tag No
Fl-1101
Fl-1 01
C10:Tag No
Fl-1201
C:Basic Setup
49:Flow User Span
10:Tag No
11:Flow Damping
108.0
FR
l/min
: : :
100
Fl-1201 TAG
0
SET
DEC
/SHIFT
INC
SET
SET
F0506-2.ai
Major item
Current setting value
Current setting value
Modified value
Sub-item
selection (C)
Major item
Sub-item
selection (C)
Indicates selection of Tag No display for B42: Display Select3.
Sub-item
Parameter
Search Mode
Parameter
Replacement Mode
Sub-item
Parameter
Search Mode
X
4
Upon selection of C: Basic Setup, the cursor will
be positioned at C10: Tag No. (Sub-item selection
screen (C))
Touch the
SET
switch to access Parameter
Replacement Mode.
Thecursorwillashonandoffontheleftofthetag
number. As “FI-1201” is to be setup in this example,
touch the
DEC /SHIFT
switch to move the cursor
to the position for multiples of 100.
At the position for multiples of 100, touch the
INC
switch to change the “1” to “2.” When the
setting has been changed to “FI-1201,” touch the
SET
switch.
When the
SET
switch is touched, the entire
displayashesonandoff.Conrmthatthesetting
has been correctly changed to “FI-1201,” and then
xthissettingbytouchingthe
SET
switch once
again.
NOTE
When no operations are carried out for 20
secondsintheashingstate,thesystemwill
automatically return to the Sub-item Parameter
Search Mode.
When the operations except
SET
are carried
out, the parameter cannot be set.
The system automatically returns to sub-item
selection screen (C).
After returning to Display Mode by touching using
the (
ESC
) twice as escape switches, it will be
possibletoconrmthemodiedcontent(ifDisplay
Select has been setup to display the tag number.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXR and is the answer not in the manual?

YOKOGAWA ADMAG AXR Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals