12-13
IM 01E30D01-01EN
<12. OUTLINE>
12.6 OPTIONAL SPECIFICATIONS
○
Available – : Not available
Item Specications
Applicable
Model
Code
General
Explosion
proof
AXR***G
AXR***C
Lightning
Protector
A lightning protector is built into the power supply and digital output terminals.
Supply voltage: 14.7 to 32 V DC
Allowablecurrent:Max600A(8/20μs)
○ ○
A
Down-scale of
output status at
alarms (Burnout)
The current output is set to 3.2 mA (-5%) or less during System Alarms (hardware
failure), Process Alarms and Setting Alarms.
Standard products are delivered with a setting of 21.6 mA (110%) or more during
System Alarms, Process Alarms and Setting Alarms.
Note: In the case of style: S1 with optional code C1, the current output is set up to 3.2
mA or less during System Alarms, and 3.8 mA during Process Alarms and Setting
Alarms.
○ ○
C1
Mass Unit
Setting
Theowratespan,transmissionpulseweight,andtotalizerdisplaypulseweightcan
besetintermsofmassunit.Specifythedensityoftheprocessuidwhenordering
inadditiontothemassowratespan,transmissionpulseweight(formassunit),and
totalizer display pulse weight (for mass unit).
1. Density
a. Available density Numerics:
Specify the numeric within the value of 500 to 2000 kg/m
3
, 4.2 to 16.7 lb/gal,
or31.2to124.8lb/cf.Anditcanbeuptovedigits,toamaximumof32000
ignoring the decimal point. A fraction is limited to the fourth decimal place.
b. Available density units: kg/m
3
, lb/gal, lb/cf
Example: A water density is about 1000kg/m
3
. In this case specify “1000kg/m
3
”.
However a density is changed by temperature.
Specify the actual density.
(The 1000kg/m
3
is equivalent to 8.345lb/gal and 62.43lb/cf.)
2.Themassowratespan,transmissionpulseweight,andtotalizerdisplaypulse
weight
a. Available density Numerics:
Specify the numeric within the value of 0.0001 to 32000. And it can be up to
vedigits,toamaximumof32000ignoringthedecimalpoint.Afractionis
limited to the fourth decimal place.
b. Mass Units Available mass units: t, kg, g, klb, lb
Available time units: /d, /h, /min, /s
Note1:Inthecaseofspecifyingthemassowspan,calculatethevolumetricowspan
bythesettingdensity,andspecifytheavailablevalueinthemassowspan.
Note2: In the case of transmission pulse weight and totalizer display pulse weight,
specifythemassunitwhichwasspeciedastheowunit.
○ ○
MU
Waterproof
Glands
Waterproof glands are attached to the electrical connections.
Available only for JIS G1/2 female electric connections.
○
– EG
Waterproof
Glands with
Union Joints
Waterproof glands with union joints are attached to the electrical connections. Avail-
able only for JIS G1/2 female electric connections.
○
– EU
Plastic Glands
Plastic glands are attached to the electrical connections.
Available only for JIS G1/2 female electric connections.
○
– EP
Mirror Finished
PFA Lining
MirrornishingonthePFAlininginsideofthetubetothesmoothnesslining.
The Ra is average of measured values on several points.
Size150to200mm(6to8in):Ra0.05to0.15μm
MirrornishedPFAliningisstandardforsize25to100mm(1to4in.)andoptional
forsize150to200mm(6to8in.)speciedbyoptionalcodePM.
○ ○
PM
Stainless Steel
Tag Plate
A pendant tag plate of JIS SUS304 (AISI 304 SS/EN 1.4301 equivalent) is provided.
Choose this option when a pendant tag plate is required in addition to the standard
nameplate with the tag number inscribed on it.
Dimension (Height × Width): Approx. 12.5 (4.92) × 40 (15.7) mm (inch)
○ ○
SCT
Direction change
of the electrical
connection (*1)
+90 degrees rotated converter to change the direction of the electrical connection.
○ ○
RA
+180 degrees rotated converter to change the direction of the electrical connection.
○ ○
RB
–90 degrees rotated converter to change the direction of the electrical connection.
○ ○
RC